-
In the lateral flow assay, the liquid moves forward after loading, and the labeled antibody and antigens form a sandwich complex with the capture antibody. The remaining labeled P24 antibody binds to the control line coated with antibody goat anti-mouse to form a complex. The peak area of the detection line increased with the increase of P24 antigen concentration, and the peak area of the quality control line decreased with the increase of P24 antigen concentration. Corresponding test results show that the P24 antigen concentration is the abscissa and the T/C peak area ratio is the ordinate. When the sample is negative, the fluorescence value of the detection line is low or not detected at all; when it is positive, the higher the P24 concentration, the higher the fluorescence value of the detection line. Schematic diagram of fluorescent microsphere immunochromatography at different concentrations of P24 antigen is shown in Figure 3. Whether results are positive or negative, the control line can always display the fluorescence value. If the control line has no fluorescence value, the results are invalid regardless of the fluorescence intensity of the detection line.
-
We investigated the effects of changing the order of addition of EDC and the monoclonal antibody. One option is to first add the antibody to the fluorescent microsphere solution, and then add the EDC to the solution after the antibody aggregates near the microspheres. A second option is to add EDC activated fluorescent microspheres COOH, centrifuge to remove excess EDC solution and resuspend, and then add antibodies to couple the fluorescent microspheres and antibodies. After labeling, diluting the labeled fluorescent microspheres 50 times using stock buffer (10 mmol/L Tris.Cl, 1% NaCl, and 2% BSA), we added 2 μL of the 1:50 diluted fluorescent microspheres to 50 μL chromatography buffer (10 mmol/L Tris.Cl, 0.2% Triton X-100, 1% Tween-20) to prepare an antibody working solution. We diluted P24 antigen to a concentration of 0, 1, 5, or 10 ng/mL. After mixing the working solution and different concentrations of antigen to 102 μL, we pipetted to add all the liquid to the P24 test card well and allowed it to stand at room temperature for 15 min. Following this, we inserted the test card into the portable fluorescent scanner KY-100, read the T/C ratio, and recorded the result. Each experiment was repeated three times to determine the optimal order of addition; results are shown in Figure 4. Under the same conditions, the second method had better labeling efficiency and thus was chosen for further analysis.
-
We used HCl and NaOH to change the pH of the HEPES buffer to 6.5, 7.0, 7.5, and 8.0, respectively, and then added 10 μL of 5% fluorescent microspheres to 100 μL of HEPES buffer at different pH values and labeled them according to the antibody linked to fluorescent microspheres method. Then, we stored them in the corresponding pH storage solution. The fluorescent microspheres labeled under conditions of pH 6.5 and pH 8.0 showed baseline drift in detection, so the fluorescent microspheres at pH 7.0 and pH 7.5 were mainly detected. The fluorescent microspheres labeled under conditions of pH 7.0 and pH 7.5 were detected at intervals for a total of 50 days (Figure 5). With the increase in storage time, the detection value of the fluorescent microsphere negative control at pH 7.0 increased slowly and then increased sharply at day 50. At pH 7.5, the negative and positive controls had relatively constant values, indicating that the fluorescent microspheres are relatively stable at this pH. Therefore, HEPES buffer at pH 7.5 was selected as the reaction solution.
-
We labeled 100 μL of 0.5% fluorescent microspheres with 30, 50, 70, 90, and 110 μL of monoclonal antibody at a concentration of 0.5 mg/mL. That is,the mass ratios of the antibody to the fluorescent microspheres were 30, 50, 70, 90, and 110 μg/mg. Under the same conditions, the mass ratio of antibody to fluorescent microspheres was 90 μg/mg, which had better labeling efficiency (Figure 6). Therefore, we decided to add 90 μL of antibody at 0.5 mg/mL.
-
The capture antibody was sprayedon the nitrocellulose membrane at concentrations of 0.5, 1.0, 1.5, 2.0, 2.5, and 3.0 mg/mL, and the fluorescent microspheres were labeled using the optimal labeling conditions described above. Detection of positive samples with a P24 antigen concentration of 1 ng/mL and negative samples, respectively. The concentration of the nitrocellulose membrane-immobilized antibody was selected based on the concentration of the largest peak area ratio of the positive sample and the smallest area ratio of the negative sample. The results are shown in Figure 7. Under the same conditions, an antibody concentration of the test line of 2 mg/mL, had the best detection efficiency. Therefore, we selected this concentration.
-
When the amount of the sample is large, components in the serum affect antigen-antibody binding, so selecting a suitable amount of the sample is important. Diluting the labeled fluorescent microspheres 50 times, we added 2 μL of the antibody to 50, 60, 70, 80, or 90 μL diluent to prepare an antibody working solution. After diluting the P24 antigen to 1 ng/mL with negative serum, we added 10, 20, 30, 40, or 50 μL of antigen to 92, 82, 72, 62, or 52 μL of the working solution to maintain a total volume of 102 μL. Detection of positive samples with a P24 antigen concentration of 1 ng/mL and negative samples, respectively. The sample addition amount was selected based on the concentration of the largest peak area ratio of the positive sample and the smallest area ratio of the negative sample. The results are shown in Figure 8. Under the same conditions, a sample addition amount of 20 μL had the best detection efficiency. Therefore, we selected this amount.
-
The time of the immunochromatographic reaction is a key factor in this experiment. If the reaction time is too short, the antigen-antibody binding will be insufficient. If the reaction time is too long, the advantage of short detection time will be lost. Therefore, it is important to select a suitable reaction time for the development of test strips. We recorded the peak area ratio every 2.5 min. Negative controls were tested in the same manner. Each experiment was repeated three times, and we selected a time when the peak area ratio was normal and the length was shorter as the reaction time; results are shown in Figure 9. At 10 min, the detection value of the negative control tended to be stable, and the value of the positive control continued to increase with time. To satisfy the conditions for rapid detection, 15 min was selected as the reaction duration.
-
After the fluorescent microsphere-labeled antibody (90 μg/mg) was diluted 50 times, we added 2 μL of the antibody to 80 μL of diluent to prepare an antibody working solution. After diluting the P24 standard with negative serum, the concentrations of P24 antigen were 0.0075, 0.015, 0.03125, 0.0625, 0.125, 0.25, 0.5, 1, 1.25, 2, 2.5, 4, 5, 8, and 10 ng/mL. We added 20 μL of P24 standard to the working solution at different concentrations. After thorough mixing, we pipetted to add all the liquid to the P24 test card well, and allowed it to stand at room temperature for 15 min. After that, we inserted the test card into the portable fluorescent scanner KY-100, read the ratio, and recorded the result. Each concentration point was detected in parallel three times and was negative. The control was tested 10 times in parallel. The P24 standard concentration was plotted on the abscissa and the ratio of peak area is plotted on the ordinate; results are shown in Figure 10. When the P24 concentration was low or high, there were different performance trends, so the curve is fitted in sections. When the P24 antigen concentration is low, the range is 7.5 pg/mL – 1 ng/mL, the fitting function is y = 1.566x2 + 5.451x + 0.0863, and R2 is 0.9969. When the concentration of P24 antigen is high, the range is 1–10 ng/mL, the fitting function is y = 0.07715x2 + 5.607x + 0.5762, and R2 is 0.9796.
-
The LOD is the minimum detectable amount of the method, which is the concentration corresponding to the mean of the negative control fluorescence signal ratio(
$\bar{x}$ ) plus 3 times the standard deviation (s) on the working curve. After substituting the calculated value into a working curve of 7.5 pg/mL to 1 ng/mL, the LOD of the method was determined to be 3.4 pg/mL. -
The concentration of P24 in serum was determined by fluorescent microsphere immunochromatography. The results of recovery and precision is shown in Table 1. The intra-assay recoveries of the low, medium, and high groups of samples were 92.0%, 97.8%, and 99.6%, respectively, and the average recovery was 96.5%. The inter-assay recoveries were 88.0%, 105%, and 98.9%, and the average recovery was 97.3%. The intra- and inter-assay CV values of the low, medium, and high concentrations were 5.4%–8.6%, and 8.5%–11.0%, respectively. These values meet production requirements which the intra-assay CV value must be less than 10% and the inter-assay CV value must be less than 15%.
P24 antigen addition concentration (ng/mL) Intra -assay Inter-assay Average value (ng/mL) Recovery (%) Standard deviation CV (%) Average value (ng/mL) Recovery (%) Standard deviation CV (%) 0.05 0.046 92.0 0.03 8.6 0.044 88.0 0.03 10.5 1.00 0.978 97.8 0.38 5.4 1.050 105.0 0.77 11.0 5.00 4.981 99.6 1.92 6.5 4.943 98.9 2.53 8.5 Table 1. Three spiked concentrations (0.05, 1, and 5 ng/mL) of P24 were analyzed for intra- andinter-assay recovery and precision studies
-
Sixty-six clinical serum samples were tested, and a comparison of fluorescent microsphere immunochromatographic test strips and Hebei Medical University P24 antigen test kits was performed (Figure 11). Since the quantitative concentration of the Hebei Medical Science P24 antigen detection kit is 0–80 pg/mL, data within the detection range were compared for mapping. For samples outside the quantitative range of the kit, methodological comparisons were performed by comparing the detection rates. The comparison of detection rates is shown in Table 2. It can be seen from Figure 11 that the function is y = 0.8019x + 3.186, R2 = 0.9393, showing that the fluorescent microsphere immunochromatographic test strip is highly correlated with the ELISA method for determining the concentration of P24 in AIDS human serum, and has good consistency.
Type of samples Number of samples ELISA Fluorescent microsphere immunochromatographic assay Number of samples detected Detection rate (%) Number of samples detected Detection rate (%) Positive samples 47 28 59.6 33 70.2 Negative samples 19 18 94.7 19 100.0 Table 2. Comparison of fluorescent microsphere immunochromatographic assay and ELISA
A New Method for Ultra-sensitive P24 Antigen Assay Based on Near-infrared Fluorescent Microsphere Immunochromatography
doi: 10.3967/bes2020.024
- Received Date: 2019-09-29
- Accepted Date: 2020-02-21
-
Key words:
- Fluorescent microsphere immunochromatography /
- HIV /
- P24 antigen /
- POCT
Abstract:
Citation: | WANG Qi, HOU Mei Ling, LIU Li Peng, MA Jing, ZHANG Xiao Guang, ZHOU Zhi Xiang, CAO Yu Xi. A New Method for Ultra-sensitive P24 Antigen Assay Based on Near-infrared Fluorescent Microsphere Immunochromatography[J]. Biomedical and Environmental Sciences, 2020, 33(3): 174-182. doi: 10.3967/bes2020.024 |